# pulse-distribution-amplifier
|
|
|
|
|
|
## Current version PDA 2017.07
|
|
|
## Rise-time (PDA 2017.07)
|
|
|
|
|
|
Rise-time measurement on 2019-12-16. BNC-cable to 4 GHz LeCroy oscilloscope, DC 50 Ohm termination, 10%-to-90% rise time measurement result is <700 ps.
|
|
|
Rise-time measurement on 2019-12-16. BNC-cable to 4 GHz LeCroy oscilloscope, DC 50 Ohm termination, 10%-to-90% rise time measurement result is <700 ps. The 1PPS output uses five parallel outputs of the clock-buffer, driving about 2.7V into a 50 Ohm load and 3.3V into 1 MOhm.
|
|
|
|
|
|
![pda_rise-time_2019-12-16](uploads/aea60c5d88cfe09efa10559d16d23e45/pda_rise-time_2019-12-16.png)
|
|
|
|
|
|
## Ouput skew
|
|
|
|
|
|
Output-to-output time difference measured with Keysight 53230A counter. Measurements done before any tuning with the trimmer-capacitors. After tuning it is possible to achieve <10 ps output-to-output skew. The delay from the input to any output has not been measured accurately, but it should be around 15 ns (to be verified).
|
|
|
|
|
|
![pda_output_skew_2021-05](uploads/8bf10af50e902a8aa98fae175f451db7/pda_output_skew_2021-05.png)
|
|
|
|
|
|
## Past versions
|
|
|
|
... | ... | |