... | ... | @@ -43,8 +43,8 @@ la Lune** |
|
|
with a source spectrum of natural unfiltered sunlight.
|
|
|
- Any value between 50.0 mMED/hr and 199.9 mMED/hr may be used to
|
|
|
calibrate.
|
|
|
- Difference between the two meters may be +/- 0.3 mMED/hr.
|
|
|
- Verify with no light on the sensor a reading of 00.0 +/- 0.3
|
|
|
- Difference between the two meters may be ± 0.3 mMED/hr.
|
|
|
- Verify with no light on the sensor a reading of 00.0, ± 0.3
|
|
|
mMED/hr.
|
|
|
3. Disable the digit on the right with an internal switch or jumper.
|
|
|
|
... | ... | |