Design
|
|
Design
|
|
|
|
|
... | @@ -58,18 +61,21 @@ Design |
... | @@ -58,18 +61,21 @@ Design |
|
|
|
|
|
##### Calculation of required sensitivity and sensor
|
|
##### Calculation of required sensitivity and sensor
|
|
|
|
|
|
- 1 MED/hr = 5.8 µW/cm² (from [Units](/uv-meters#units))
|
|
- 1 MED/hr = 5.8 µW/cm² (from [Units](/uv-meters#units)).
|
|
- Full display 199 mMED/hr = 0.199 MED/hr = 1,1542 µW/cm² = 1.999 V
|
|
- Multiply by correction factor of 0.24 for a wideband TOCON ABC (that
|
|
- 5V TOCON should have max. irradiance at 5/1.999 \* 1.1542 = 2.89
|
|
is more sensitive for UVA) to correct so that under natural sunlight
|
|
µW/cm²
|
|
gives same reading in med/hr as a perfect erythema meter.
|
|
|
|
- See [*Comparison of different UV sensors for use in an XP
|
|
|
|
radiometer*](https://www.ohwr.org/project/uv-mask/wikis/Documents/Comparison-of-dierent-UV-sensors), A.
|
|
|
|
Butterworth, December 16, 2018 (paragraph above Table 1).
|
|
|
|
- Full display 199 mMED/hr = 0.199 MED/hr = (1,1542\*0.24) µW/cm² =
|
|
|
|
1.999 V
|
|
|
|
- 5V TOCON should have max. irradiance at (5/1.999) \* 1.1542\*024
|
|
|
|
= 0.693 µW/cm²
|
|
- [TOCON\_ABC3](https://sglux.de/en/product/tocon_abc3-en/)
|
|
- [TOCON\_ABC3](https://sglux.de/en/product/tocon_abc3-en/)
|
|
max. irradiance (saturation limit) at 290 nm (peak
|
|
max. irradiance (saturation limit) at 290 nm (peak
|
|
wavelength) is approx. 1.8 µW/cm2
|
|
wavelength) is approx. 1.8 µW/cm2
|
|
Not sensitive enough.
|
|
Need attenuation of around a factor of 3, handling tolerance
|
|
- [TOCON\_ABC4](https://sglux.de/en/product/tocon_abc4-en/)
|
|
|
|
max. irradiance (saturation limit) at 290 nm (peak
|
|
|
|
wavelength) is approx. 18 µW/cm2
|
|
|
|
Need attenuation of around a factor of 6, handling tolerance
|
|
|
|
of ±20%.
|
|
of ±20%.
|
|
|
|
|
|
Likely not needed: digital calibration could be made with a [TOCON
|
|
Likely not needed: digital calibration could be made with a [TOCON
|
... | @@ -131,5 +137,5 @@ Other |
... | @@ -131,5 +137,5 @@ Other |
|
|
|
|
|
-----
|
|
-----
|
|
|
|
|
|
Erik van der Bij - 7 January 2019
|
|
Erik van der Bij - 10 January 2019
|
|
|
|
|