- The tolerance of the TOCONs is ±20%, some of them may not give
|
|
- The tolerance of the TOCONs is ±20%, some of them may not give
|
|
us the required output: 2.3V\*0.8= 1.84V instead of 2.000V.
|
|
us the required output: 2.3V\*0.8= 1.84V instead of 2.000V.
|
|
Should take this in account that 1.84V can be calibrated to read
|
|
Should take this in account that 1.84V can be calibrated to read
|
|
as 2.00V.
|
|
as 2.00V. 2.3V\*1.2=2.76V. So calibration range must be 1.8V -
|
|
|
|
2.8V.
|
|
- For comparison 0.2 MED/hr should give 0.79 V on the TOCON E1.
|
|
- For comparison 0.2 MED/hr should give 0.79 V on the TOCON E1.
|
|
|
|
Calibration range 0.6V - 1V.
|
|
- Assuming the powers that are quoted by sglux are in W/cm² are
|
|
- Assuming the powers that are quoted by sglux are in W/cm² are
|
|
normalised to the sensitivity spectrum of the sensor.
|
|
normalised to the sensitivity spectrum of the sensor.
|
|
|
|
|