Skip to content

  • Projects
  • Groups
  • Snippets
  • Help
    • Loading...
  • Sign in
F
FMC ADC 100M 14b 4cha
  • Project
    • Project
    • Details
    • Activity
    • Cycle Analytics
  • Repository
    • Repository
    • Files
    • Commits
    • Branches
    • Tags
    • Contributors
    • Graph
    • Compare
    • Charts
  • Issues 0
    • Issues 0
    • List
    • Board
    • Labels
    • Milestones
  • Merge Requests 0
    • Merge Requests 0
  • Wiki
    • Wiki
  • image/svg+xml
    Discourse
    • Discourse
  • Members
    • Members
  • Collapse sidebar
  • Activity
  • Graph
  • Charts
  • Create a new issue
  • Commits
  • Issue Boards
  • Projects
  • FMC ADC 100M 14b 4cha
  • Issues
  • #10

Closed
Open
Opened Jun 02, 2021 by scott cogan@cogansco
  • Report abuse
  • New issue
Report abuse New issue

Document 4th input voltage range, +/-1V?

While experimenting with the v6 FMC and my application, I found that the +/-0.5V range was not quite enough voltage range for me, but the +/-5V was too much, as noise was too high. I found an undocumented +/-1.0V (2Vpp) range

Below is a table of which switches would be enabled for each of the 3 documented ranges, and I found a 4th range using switch settings = 0x64 which implements a +/-1V, or 2Vpp range. The noise level is the same as 1Vpp range, but has 2X the range, so SNR is 6 dB better than 1Vpp. There did not appear to be any change to the bandwidth or any other compromise. I submit this as an issue ticket for consideration in case the documentation could be updated, or perhaps there is a good reason why this switch configuration was not considered previously. Comments welcome. image image